Section 1 Specifications
Specifications are subject to change without notice.
Installation Environment
Bench Space Needed 1.5 m (4 ft)
Total Height, tubes and block 33 cm (13 inches)
Ambient Operating
Temperature Range
5–40 °C
Altitude Up to 2 Km (6800 ft) above sea level
Relative Humidity Up to 92.5%, non-condensing
Shelf Weight 20 kg (45 lb)
Ventilation Needed
Ordinary air-conditioned lab. Provide at least 6 inches distance between the Micro Dist
block and all other vertical surfaces.
Safety Laboratory 61010-1, 61010-2-010 (UL, CSA, EN Safety Standards)
EN61010 Pollution Degree II
EN61010 Installation Category II
Micro Dist Heating Block
AC Input Frequency
100–115 V, 50/60 Hz
230 V, 50/60 Hz
Maximum Active Current (A)
Rating
100 V, 9 A
115 V, 11 A
230 V, 5 A
Fuse Rating
100–115 VAC Model: 15 A slow-blow, 125 VAC
230 VAC Model: 6.3 A slow-blow, 250 VAC
Block Operating
Temperature Range
120 °C typical; 30–135 °C, L-type thermocouple, PID variable cycle controller
Controller Setpoint Error <1% of setpoint temperature
Controller Precision 0.5 °C
Hole-to-Hole Temperature
Variation
<0.5 °C
Sample Capacity 21 Micro Dist™ Tubes
Micro Dist Tubes
Sample Size 6.0 mL capacity
Tube Material Analytical-grade pure polypropylene
Compliance
Certifications
100–115 V model: approved to UL and CSA safety standards (cETLus mark);
230 V model: CE certified by Hach