EasyManuals Logo

Hach Micro Dist User Manual

Hach Micro Dist
70 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #5 background imageLoading...
Page #5 background image
5
Section 1 Specifications
Specifications are subject to change without notice.
Installation Environment
Bench Space Needed 1.5 m (4 ft)
Total Height, tubes and block 33 cm (13 inches)
Ambient Operating
Temperature Range
5–40 °C
Altitude Up to 2 Km (6800 ft) above sea level
Relative Humidity Up to 92.5%, non-condensing
Shelf Weight 20 kg (45 lb)
Ventilation Needed
Ordinary air-conditioned lab. Provide at least 6 inches distance between the Micro Dist
block and all other vertical surfaces.
Safety Laboratory 61010-1, 61010-2-010 (UL, CSA, EN Safety Standards)
EN61010 Pollution Degree II
EN61010 Installation Category II
Micro Dist Heating Block
AC Input Frequency
100–115 V, 50/60 Hz
230 V, 50/60 Hz
Maximum Active Current (A)
Rating
100 V, 9 A
115 V, 11 A
230 V, 5 A
Fuse Rating
100–115 VAC Model: 15 A slow-blow, 125 VAC
230 VAC Model: 6.3 A slow-blow, 250 VAC
Block Operating
Temperature Range
120 °C typical; 30–135 °C, L-type thermocouple, PID variable cycle controller
Controller Setpoint Error <1% of setpoint temperature
Controller Precision 0.5 °C
Hole-to-Hole Temperature
Variation
<0.5 °C
Sample Capacity 21 Micro Dist™ Tubes
Micro Dist Tubes
Sample Size 6.0 mL capacity
Tube Material Analytical-grade pure polypropylene
Compliance
Certifications
100–115 V model: approved to UL and CSA safety standards (cETLus mark);
230 V model: CE certified by Hach

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach Micro Dist and is the answer not in the manual?

Hach Micro Dist Specifications

General IconGeneral
BrandHach
ModelMicro Dist
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals