Adjust the display contrast
1. Push the menu key and select Polymetron 9500 SETUP>DISPLAY SETUP>DISPLAY
CONTRAST.
2. Use the up and down arrow keys to adjust the contrast to a value between the minimum of
+1 and the maximum of +9.
Controller configuration information
General information about configuration options is listed in the table.
1. Push the menu key and select Polymetron 9500 SETUP.
Option Description
SECURITY SETUP Sets the passcode preferences.
OUTPUT SETUP Configures the controller analog outputs.
RELAY SETUP Configures the controller relays.
DISPLAY SETUP Configures the controller display.
ADJUST ORDER—View and modify the measurement display order.
• SEE CURRENT ORDER—View the current display order
• ADD MEASUREMENTS—Add selected measurements to the display
• REMOVE MEASUREMENTS—Remove selected measurements from the display
• REORDER LIST—Select one or more measurements and change their order in the
display
• SEE DEFAULT ORDER—View the default display order
• SET TO DEFAULT—Set the display order to the default configuration
Note: Some of the above will not be available if no adjustment is possible for that
option (e.g. REORDER LIST and REMOVE MEASUREMENTS will not be available if
only one measurement is selected for display).
DISPLAY CONTRAST—Adjusts the contrast of the controller display.
EDIT NAME—Assigns a name to the controller.
SET DATE/TIME Sets the controller time and date.
DATALOG SETUP Configures data logging options. Available only if CALCULATION has been setup.
MANAGE DATA Select the device from the list of installed components and then select VIEW DATA
LOG or VIEW EVENT LOG depending on the type of log entry to view. Specify the
selection period to list all log entries matching the selection criteria. Push the up and
down arrows to select an entry and then push enter to view more details.
ERROR HOLD
MODE
HOLD OUTPUTS—Holds outputs at last known value when controller loses
communication with the sensor.
TRANSFER OUTPUTS—Switches to transfer mode when controller loses
communication with the sensor. Outputs transfer to a pre-defined value.
22 English