9. Put the sample
cell in the sample
cell holder with the
triangle on the
sample cell aligned
with the reference
mark on the sample
cell holder. Push the
lid closed until a click
is heard.
10. Push Read (or
Done if in
continuous mode).
Wait for the
instrument to read
the sample.
Note: If auto store is
off, push Options >
Store to save the data.
6.4 Data management
6.4.1 Show the recorded data
All the recorded data is kept in the data log. There are three types of data logs:
Reading log—Shows the recorded measurements.
Calibration log—Shows the calibration history.
Verification log—Shows the verification history.
1. Push Data Log and select the applicable data log.
2. To show the details of a log entry, select the log entry and then push View Details.
Note: To add a comment to the log entry, push the comments icon.
3. To show only some of the data, push Filter, then select On. The Filter Settings window opens.
4. Select an option.
Option Description
Time Interval Selects only the data that was stored during a specific time interval.
Operator ID Selects only the data that was stored with a specific operator ID.
Sample ID Selects only the data from the Reading Log that was stored with a specific
sample ID.
6.4.2 Send data to a connected device
The instrument can send data to a USB memory device or Seiko DPU-S445 printer. For best results,
use only USB 2.0 memory devices. The instrument makes a logger folder on the device and saves
the data as a .bmp, .csv or .xml file.
1. Connect a USB memory device or cable to a USB port on the instrument.
2. Connect the other end of the cable to the printer, if applicable.
3. Go to Setup>Peripherals. The connection status shows Connected. If the status shows Not
Connected, make sure to use the recommended devices.
4. Push Data Log and select the applicable log.