EasyManuals Logo

Hanna Instruments HI 9829 User Manual

Hanna Instruments HI 9829
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #17 background imageLoading...
Page #17 background image
17
3.2 PROBE SPECIFICATIONS
Non-logging Probe Logging Probe
Sample Environment Fresh, brackish, seawater
Waterproof protection IP68
Computer Interface NA USB PC (HI 76982910)
Internal Battery Type NA 4 X 1.5V Size AA Alkaline
Typical Battery Life NA See below
Memory NA 140,000 measurements
(single parameter logged)
35,000 measurements
(all parameters logged)
Operating Temperature -5 to 55° C *
Storage Temperature -20 to 70° C
Maximum Depth 20 m (66 ft.) *
Dimensions HI 7609829 342mm (13.5”), HI 7629829 442mm (17.4"),
(without cable) dia=46 mm (1.8”) dia=46 mm (1.8")
HI 7619829 382 mm (15.1”), HI 7639829 482 mm (19.0"),
dia=46 mm (1.8”) dia=46 mm (1.8")
Weight HI 7609829 570g (20.1 oz.) HI 7629829 775g (27.3 oz.)
(with batteries and sensors) HI 7619829 650g (22.9 oz.) HI 7639829 819g (28.9oz.)
Cable Specification Multistrand-multiconductor shielded cable with internal strength
member rated for 68 kg (150 lb) intermittent use
Wetted Materials Body: ABS
Threads: Nylon
Shield: ABS/ 316 SS
Temp probe: 316 SS
O-rings: EPDM
* Reduced for ISE sensors
LOGGING PROBE BATTERY LIFE
Interval All channels logging All channels logging
(no averaging) (10 sample averaging)
1 - 5 sec 72 hours 72 hours
1 min 22 days 11 days
10 min 70 days 65 days

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 9829 and is the answer not in the manual?

Hanna Instruments HI 9829 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 9829
CategoryMultimeter
LanguageEnglish

Related product manuals