EasyManua.ls Logo

Hanna Instruments HI 9829 - Probe Logging Setup

Hanna Instruments HI 9829
99 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
67
11.2.2 Continuous meter log
Select “Start meter log” to log the currently en-
abled parameters at the set logging interval on
the meter.
To set the logging interval, highlight “Start meter
log” and press <Options>. The log interval time
can set from 1 second to 3 hours. Press
<Modify> and use the arrow keys and keypad
to enter the desired log interval. Press <Accept>
to confirm.
Press <Select> to edit the lot, remark, or tag, see
section 11.4.
To stop the meter log, enter the log menu and
select <Stop meter log> .
11.3 PROBE LOG (only for logging probes)
Select “Start probe log” to start a log with the
current settings. Press <Options> to change the
log settings.
Probe Log Options
To edit the lot remark, or tag, see section 11.5.
The log interval time can be set from 1 second
to 3 hours. Press <Modify> to change the
logging interval. Press <Accept> to confirm.
“Select logging param.” to modify the parameters
to be logged.
To specify the log start time, highlight “Use start
time” and press <Enable>. Highlight “Log start
time” and press “Select”. Enter the desired time
and press <Accept> to confirm.
To specify the log stop time, highlight “Use stop
time” and press <Enable>. Highlight “Log stop
time” and press <Select>. Enter the desired time
and press <Accept> to confirm.

Table of Contents

Related product manuals