EasyManuals Logo

Hanna Instruments HI 9829 User Manual

Hanna Instruments HI 9829
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #60 background imageLoading...
Page #60 background image
60
ISE
From the “GLP” menu, select the “ISE”
option.
Data regarding the last ISE calibration will be
displayed: standards used, sensor type, time
and date of the calibration.
Use the arrow keys to scroll through the
stored data for the last 5 calibrations.
Press ESC to return to the “GLP” menu.
Notes If no ISE calibration has been performed or if calibration was cleared
using the “Restore factory calib.” option the offset and slope values
are set to default, and the message ”Factory calibration” is displayed.
Press <ESC> to return to the previous screen.
ORP
From the “GLP” menu select the “ORP” option.
Data regarding the last ORP calibration will be
displayed: calibration point, time and date.
Use the arrow keys to scroll through the stored
data for the last 5 calibrations.
Press ESC to return to the “GLP” menu.
Notes If no ORP calibration has been performed or if calibration was cleared
using the “Restore factory calib.” option the offset and slope values
are set to default, and the message ”Factory calibration” is displayed.
Press <ESC> to return to the previous screen.
Dissolved Oxygen
From the “GLP” menu select the “Dissolved oxy-
gen” option.
Data regarding the last D.O. calibration will be
displayed: calibration points, % saturation or
concentration, time and date.
Use the arrow keys to scroll throught the stored
data for the last 5 calibrations.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 9829 and is the answer not in the manual?

Hanna Instruments HI 9829 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 9829
CategoryMultimeter
LanguageEnglish

Related product manuals