7.8 TEMPERATURE CALIBRATION
The probe is factory calibrated for temperature readings. The user can perform a
single point temperature calibration or restore factory calibration. This proce-
dure requires a reference temperature measuring instrument.
Select “Temperature” from the “Calibration” menu.
Select “Calibrate temperature”.
Insert the probe in an isothermal bath with reference instrument and allow the
probe to come to thermal equilibrum.
Use the keypad to enter the known temperature and then press <Accept> to
confirm.
The stability timer will count down and the mes-
sage “Ready” and <Confirm> will be displayed.
Press <Confirm> to store the calibration point.
After confirmation, the following messages are
displayed “Storing” and “Calibration completed”.
Press <OK> to return to the “Calibration” menu.
Press <Measure> to return to the measurement
screen.
To restore the factory calibration, select the
corresponding option in the “Temperature
calib.” menu and then press <Select>.
7.9 ATMOSPHERIC PRESSURE CALIBRATION
Place HI 9829 in a wind-free area and choose “Custom pressure” to perform a
user calibration or “Restore factory calib”.
Note “Custom pressure” procedure requires
a reference barometer.
Select the “Atm. pressure” from the “Calibration”
menu.
Select the “Custom pressure” option.
Using the keypad, insert the numeric value that
agrees with the reference meter and then press
<Accept> to confirm.