EasyManua.ls Logo

Hanna Instruments HI 9829 - Page 39

Hanna Instruments HI 9829
99 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
39
7.2.3 pH Calibration Error Messages
The HI 9829 displays a series of messages if an error has occurred during
calibration.
If the meter does not accept a pH calibration point, a short message is displayed
to indicate the possible error source. The following screens are examples:
These are the available messages:
“Input out of scale”: the pH value is out of range. The pH sensor may
require replacement.
“Check sensor”: the electrode may be broken, very dirty or the user has
attempted to calibrate the same buffer value twice.
“Wrong buffer”: the displayed pH reading is too far from the selected buffer
value. This is often seen immediately after a buffer calibration has been
completed but before the pH sensor has been moved to the next buffer. Check
if the correct calibration buffer has been selected.
“Invalid temperature”: the buffer temperature is outside the acceptable range.
“Wrong buffer” / “Contaminated buffer” / “Check electrode”: the buffer is
contaminated or the sensor is broken or very dirty.
“Check sensor” / “Clean sensor”: the electrode is broken or very dirty.
“Wrong” / “Clear old calibration”: erroneous slope condition. These messages
appear if the slope difference between the current and previous calibration
exceeds the slope window (80% to 110%). Press the <Clear> softkey to cancel
the old data and continue the calibration procedure, or press ESC to quit the
pH calibration mode.

Table of Contents

Related product manuals