‘’GPS error’’ (only for models with GPS): the
communication with the internal GPS unit can-
not be established. Switch the meter off and on
again, then retry. If the problem persists, remove
the batteries, wait for 5 minutes and install them
again. If the problem persists, contact the
HANNA service center.
‘’Dead meter battery!’’: This message appears
if the meter batteries are too low to power the
meter and it will automatically turn off. Con-
nect the charger if using rechargeable C bat-
teries or replace the alkaline batteries to con-
tinue.
‘’Insert rechargeable batteries and slide battery
selector switch to the down position.’’: This mes-
sage appears when non-rechargeable alkaline
batteries are installed on the meter and/ or the
battery selector switch in is the wrong position,
and the user is attempting to charge then bat-
teries.
‘’Dead probe battery!’’: This message appears
if the logging probes batteries are not supply-
ing enough voltage to power the logging probe.
Replace the probe batteries.
‘’User data corrupted!’’: This message appears
when powering up the user data stored on meter
are corrupted. Restart the meter. If the problem
persists, contact the HANNA service center.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 9829 and is the answer not in the manual?

Hanna Instruments HI 9829 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 9829
CategoryMultimeter
LanguageEnglish

Related product manuals