Measurement & Probe Setup Menu 46
9.6. LOGGING
Options: Logging Type, Sampling Period, File Name, Log Note, Log Info, Sample ID
Logging Type
• Automatic: data is logged automatically at predefined time intervals (i.e. Sampling Period).
` A file name is automatically generated, complete with year/month/day, and log starting time
(e.g., 20220329_085101‑do_auto.CSV).
• Manual: data is logged every time
M
is tapped.
` A file name is automatically generated , complete with year/month/day e.g.
20220329_085101‑do_Test_001.
` Alternately, tap Create to name a manual log file with a custom suffix. e.g. 20230404_13570‑
do_Derry samples
• Autohold: Direct/Autohold reading mode must be configured.
Note: Manual and Autohold records are stored in the same log file, i.e. data logged on different days is
stored in the same file. Automatic records are stored separately.
Data logged with Autohold option selected, is identified by the “H” in the Notes column.
Sampling Period
Option available only with Automatic logging type selected.
Time‑interval options are selected from scrollable list.
File Name
Available only with Manual and Autohold logging type selected.
To create a file name, from Logging screen:
1. Tap Create.
2. Use the on‑screen keypad to enter a file name (maximum 13 characters).
3. Tap
to confirm.
Log Note & Log Info
Notes on measured data are saved together with logged data.