59 Maintenance
Cleaning the Smart Cap
Use a mild detergent and a soft‑bristled toothbrush to clean.
Rinse with water after cleaning and dry with a laboratory tissue.
Hydrate in purified water before use.
Note: Smart Caps need to be replaced every year.
Smart Cap Replacement
One year after a new cap installation, the message “opdo
®
Cap Expired” is displayed.
To maintain measurement accuracy, the Smart Cap replacement is mandatory.
Probe Replacement Cap Kit
The probe replacement cap kit contains: Smart Cap for opdo probe (1 pc.), sachet with silicone grease (6 g),
syringe (1 pc.), lens wipe (1 pc.), certification / instruction sheet (1 pc.)