9 Specications
Temperature*
Range −20.0 to 120.0 °C; −4.0 to 248.0 °F; 253.2 to 393.2 K
Resolution 0.1 °C / 0.1 °F / 0.1 K
Accuracy Refer to probe used
Compensation Automatic or Manual
Temperature user calibration 1 point, configurable
Salinity Compensation
Manual
› 0.00 to 45.00 psu or ppt
› 0.0 to 130.0 %
Logging
Type
Automatic
Manual
Autohold
Number of records
50 000 maximum per file
Stores at least 1 000 000 data points per user
Automatic interval
1, 2, 5, 10, 30 seconds
1, 2, 5, 10, 15, 30, 60, 120, 150, 180 minutes
Sample ID Incremental mode
Export option .CSV file format
Users Up to 9 users and the default administrator account
Connectivity
USB‑A 2 ports for keyboard and / or printer input or USB flash drive
USB‑C 1 port for PC connectivity and USB‑C type flash drive
Wi‑Fi & Ethernet
FTP
Web server Log transfer and download
Email
RS232 Connecting peripherals
Power supply DC adapter 100‑240 VAC to 24 VDC 2 A
Environment 0 - 50 °C / 32 - 122 °F / 273 - 323 K, maximum 95 % RH non-condensing
Dimensions 205×160×77 mm (8.0×6.2×3.0“)
Weight Approximately 1.2 kg (26.5 lbs.)
* Consider probe’s limits