Preparation before use
EN
DPO7000 Quick Guide Copyright © Qingdao Hantek Electronics Co., LTD
18
The probe compensation signal can only be used for probe compensation
adjustment and cannot be used for calibration.
5 Probe Compensation
When connecting the probe to any input channel for the first time, this adjustment is
required to match the probe with the input channel. Uncompensated or offset
probes can lead to measurement errors or errors.
Perform the first four steps of the previous section on "Function Check".
Compare the waveform with the figure below.
If necessary, use a non-metallic screwdriver to adjust the variable capacitance
on the probe until the waveform displayed on the screen is "Perfectly
compensated" as shown in the figure above. If necessary, repeat this step.
The adjustment method is shown in the following figure.