Preparation before use 
    EN 
DPO7000 Quick Guide                                                                                   Copyright © Qingdao Hantek Electronics Co., LTD 
18 
The probe compensation signal can only be used for probe compensation   
adjustment and cannot be used for calibration. 
5、  Probe Compensation 
When connecting the probe to any input channel for the first time, this adjustment is 
required to match the probe with the input channel. Uncompensated or offset 
probes can lead to measurement errors or errors. 
⚫  Perform the first four steps of the previous section on "Function Check". 
⚫  Compare the waveform with the figure below. 
 
⚫  If necessary, use a non-metallic screwdriver to adjust the variable capacitance 
on the probe until the waveform displayed on the screen is "Perfectly 
compensated" as shown in the figure above. If necessary, repeat this step. 
The adjustment method is shown in the following figure.