4
PME−MP60DP
HBM A0623-6.1 en/de/fr
1 Introduction
This Operating Manual describes only those functions which differ from the
MP60. The features of the MP60DP correspond to those of the MP60.
The MP60DP carrier−frequency measurement module has been expanded to
include a Profibus interface. The features on the CAN−interface remain the
same; the object directory is expanded to include some parameters for the
Profibus connection.
The Profibus connection is made using a 9−pin sub−D connector (conforming
to standard) on the front panel next to the transducer port.
DP protocol is used on the Profibus.
The following are communicated:
− the measured values (gross, net, peak values)
− the status of the limit switches
− control bits for taring, zeroing, peak value store control and changing the
parameter set, and
− optionally, the limit values