Operation manual
Warning
When using the hard patient plate and single soft patient plate in the operation, the medical
staff must check whether the patient plate is attached evenly to the patient skin, so as to
prevent the patient plate falling off and the burning accident of the patients.
Dual Soft Patient Plate Using
Insert plug of patient plate cord into the desired place, and clip the clamp on the end of
the cord with the dual soft patient plate. Then the patient plate is evenly attached to the
patient skin.
The LED for dual patient plate mode will illuminate, and the green stick area display
will illuminate. The number of bars illuminating depends on impedance of patient’s body
and stick area between skin and plate. Normally, 2~9 bars will light. After the patient plate
is fully attached to patient skin, the output of cut or coag mode can be activated, and at this
time the data of stick area will be memorized in the machine. If the plate is loose or
disconnected from patient skin causing the resistance value of contacting area to be
increased by 30% when operation is going, the machine will activate alarm and stop output.
Operator must re-attach the plate properly to patient skin so as to go on operating.
Power On
Turn on the power switch, and the machine type and version number will be displayed.
The machine goes into the last using setting immediately (i.e. mode setting and power
setting).