Operation manual
Safety Instructions
● High Frequency Surgical Unit should only be operated by qualified and trained
medical staff. Before operating the unit, please read the manual carefully to avoid
any operating error.
● Please use a 3-hole socket that is properly grounded for power supply, so as to make
sure that ground points of the outer metal shell, protection or function for the unit are
connected to the ground to prevent risk of electric shocking.
● If performing the surgical operation near the chest or head, the ignitable
anesthetic
drug or
oxidable gas such as nitrous oxide (N
2
O) and oxygen can not be used unless
the reagents are removed before the operation. If possible, the flame retardant reagent
can be used in cleaning and disinfection. Before using High Frequency Surgical Unit,
the ignitable reagent for cleaning and disinfection or as adhesive solvent should be
vaporized. Please pay attention to the explosive gas or fluid that may exist in the
holes of the patients’ body (such as vagina) and get rid of these explosive elements.
In addition, the gas in the body should be paid attention to because the gas may cause
fire. Some materials (such as cotton, cloth and gauze) may be ignited by the High
Frequency Surgical Unit when they are filled with oxygen.
● Please check the accessories before the operation:
ACheck the connection between accessories and the machine and the connection
between soft patient plate cord and soft patient plate.
BCheck the insulation properties of the accessories and check the unit and cable
for damage.
Do not operate on the patient with heart pacemaker or other implantation materials,
because the unit may cause disorder or stop of the heart pacemaker. If having any
questions, we should follow the confirmed proper advice.
This unit can not be acted on the heart directly.
● Prevent the contact between the patients’ skin (e.g. the contact between the arm and
body), for instance, we can put the dry gauze between them.
Patient plate must be connected and installed correctly and the contact area should be
large enough. This problem will be described in detail in the following contents.
The output power of the unit must be on the proper level for best operation effect. Do
not increase the output casually. The limitation of the output should be according to
the operation effect. In the general monopolar unit operation, the level of the output
power should be on or about 30-70W, while special operation may need higher
output, but generally not beyond 200W. If the required setting output of a special
operation is much higher than general operation, please check the installation of the
patient plate, the condition of the patient plate and the cord, the machine state, and