List of common failure
List of external failure
Phenomenon
Cause
Solution
Turn on the power switch,
showing nothing
No alternating voltage
available in the outlet
Fuse loose or broken
Check the power supply
Tighten or replace the fuse
(If broken again after tightening, it is
problem inside.)
Machine alarm
Plate plug is not properly
inserted
Plate clamp does not secure
the conductor
Secure the plate plug and replace
the plate
Secure the plate conductor
Failure to activate the
machine by hand control
pencil
Improper contact of hand
control pencil
Replace the hand control pencil
Failure to activate the
machine by pedal switch
Improper contact of pedal
switch
Pedal switch not properly
plugged
Replace the pedal switch
Securely plug
Dual plate fails to
automatically changeover
Plate plug damaged
Replace the plate wiring
Insert the hand control
pencil and the machine
startup
Hand control pencil is placed
at normally-closed status.
Replace the hand control pencil
Output electrode and cable
overheated or output
power greatly reduced
Cable broken somewhere or
plug-in unsecured
Replace the attachment or secure
the plug-in
Casing charged
Ground wire of power cable is
not correctly grounded or power
supply wire in the machine is
detached.
Correctly connect the ground wire
of power supply cable or secure the
power conducting wire in the
machine.
Heal Force EB03 36