List of common failure
List of external failure
Phenomenon
Cause
Solution
Turn on the power switch,
showing nothing
◆
No alternating voltage
available in the outlet
◆ Fuse loose or broken
◇
Check the power supply
◇ Tighten or replace the fuse
(If broken again after tightening, it is
problem inside.)
Machine alarm
◆
Plate plug is not properly
inserted
◆ Plate clamp does not secure
the conductor
◇
Secure the plate plug and replace
the plate
◇ Secure the plate conductor
Failure to activate the
machine by hand control
pencil
◆
Improper contact of hand
control pencil
◇
Replace the hand control pencil
Failure to activate the
machine by pedal switch
◆
Improper contact of pedal
switch
◆Pedal switch not properly
plugged
◇
Replace the pedal switch
◇ Securely plug
Dual plate fails to
automatically changeover
◆
Plate plug damaged
◇
Replace the plate wiring
Insert the hand control
pencil and the machine
startup
◆
Hand control pencil is placed
at normally-closed status.
◇
Replace the hand control pencil
Output electrode and cable
overheated or output
power greatly reduced
◆
Cable broken somewhere or
plug-in unsecured
◇
Replace the attachment or secure
the plug-in
Casing charged
◆
Ground wire of power cable is
not correctly grounded or power
supply wire in the machine is
detached.
◇
Correctly connect the ground wire
of power supply cable or secure the
power conducting wire in the
machine.