Operation manual
Disassembly & Installation
Disassembly
Take out EB03 surgical unit with care; keep the hard board and materials used for
package so that these could be used for the next transportation or rework.
If there is any fault or question, please do not open the unit to repair by yourself; you
should contact the manufacturer or local representative immediately.
Working Power Supply
Power supply voltage please refer to the nameplate
Working Environment
Temperature: 10-40℃
Humidity:≤80%RH
Atmospheric pressure: 700hPa~1060hPa
Keep away from any corrosive & ignitable matter.
Placed in clean and ventilated room.
Transporting
Temperature: -40 ~ +55℃
Humidity: ≤95%RH
Atmospheric pressure 500hpa ~ 1060hpa
No vibration, no collision, and no drenching in transporting.
Package Symbol
Installation
● Remove the packing of the equipment, check all accessories for damage according to
package list.
● Insert power cord and connect its own patient plate and hand control pencil. Put a moist
soap on the patient plate.
● Turn on the unit, apply proper power (less than 50W), and activate the hand control
pencil, causing sparks when cutting soap.
● After confirming the power output, remove the accessories and deliver the complete
machine to the customers.
Heal Force EB03
4