Daily check on accessories
●Attached cable insulation layer shall not be exposed to decolour, deterioration
or damage. Leading wire shall not be visibly folded.
●Cable joint shall not be loose and detached.
●Hard plate shall be wrapped and kept intact. Fouling shall not be seen on the
plate.
Warranty commitments
We provide warranty commitments for the host machine and pedal switch.
● Warranty period may take effect since the date of procurement and last for one
year.
● For the machine under warranty period, we will not take responsibility for the
failure due to the misuse by the customer.
● Within the warranty period, the customer shall not uncover the host machine
or disassemble the pedal switch. Otherwise, the warranty may become null
and void.
●We will charge the repair beyond the warranty period.
●Provide the wiring diagram and list of components to the approved
maintenance unit and people.
Troubleshooting
Since the whole electrotome system is quite complex, you shall abide by the
following steps to do troubleshooting:
1. Assess if misoperation or equipment failure
For example: soft plate being used is not secured on the patient, leading to
failure to activate the alarm.
2. Assess if host machine failure or damage to accessory
For example: failure to startup may somehow attribute to the hand control
pencil switch or pedal switch.
3. Assess if problem outside the machine or inside the machine; The next section
will give the details.
4. Assess that the exact part of machine inside fails.
Troubleshooting on the failure inside the machine require the training.
Regular check
The user shall carry out some necessary check on performance and safety on a
regular basis.