EasyManua.ls Logo

Heal Force EB03 - Safety Indices; Classification of Product

Default Icon
37 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation manual
Safety indices
The machine is manufactured in accordance with the state safety indices
standard.
Neutral electrode is separated from the ground in case of high frequency.
High-frequency drain current <150mA
Low-frequency drain current: to ground <0.5mA
Casing<0.1mA
Patient<0.01mA
Classification of product
Category of equipment: Category I
Type of equipment: Type CF
Power supply voltage please refer to the nameplate
Rated input current of equipment: 5A
Application appliances available with the equipment: output circuit for
surgical electrode and neutral electrode
Equipment’s protection from discharge effect of defibrillator
Classified as common equipment
Operation mode of equipment: Continuous operation with loading in
intermittent pattern
Portable equipment
Heal Force EB03 29