Operation manual
Safety indices
The machine is manufactured in accordance with the state safety indices
standard.
Neutral electrode is separated from the ground in case of high frequency.
High-frequency drain current <150mA
Low-frequency drain current: to ground <0.5mA
Casing<0.1mA
Patient<0.01mA
Classification of product
● Category of equipment: Category I
● Type of equipment: Type CF
● Power supply voltage please refer to the nameplate
● Rated input current of equipment: 5A
● Application appliances available with the equipment: output circuit for
surgical electrode and neutral electrode
● Equipment’s protection from discharge effect of defibrillator
● Classified as common equipment
● Operation mode of equipment: Continuous operation with loading in
intermittent pattern
● Portable equipment
Heal Force EB03 29