Output parameters
Max.
power
(W)
Rated
load (Ω)
Duty
ratio
Max. open circuit voltage
peak value (V)
Pure cut 350 500
100%
1800
Blend 1 300 500
68%
2700
Blend 2 200 500
56%
2400
Blend 3 150 500
24%
2900
Soft coag 100 500
16%
3600
Spot coag 100 500
16%
4600
Bipolar coag 50 100
100%
240