Operation manual
Operations During the Operation Process
Mode Setting
Monopolar cutting
Work mode
Power (watt)
Pure cut
0-350w
Blend 1
0-300w
Blend 2
0-200w
Blend 3
0-150w
Monopolar coagulation
Soft
0-100w
Point
0-100w
Bipolar coagulation
Standard
0-50w
The working mode will be selected according to operation requirements by each mode selecting
switch, and the LED corresponded to the mode will illuminate.
Power Adjustment
After performing the mode setting, press cut or coag power tuning switch to
set output power and the set value will be displayed on the display.
Pressing the power tuning switch for more than 3 seconds, the power setting
can be accelerated
.
Note: When using this unit for the first time, you should be sure to start from lower power.
Output Startup
The cut and coag output are respectively controlled by cut and coag switch of
the hand control pencil or cut and coag pedal of pedal switch. The yellow
switch and yellow pedal control cut, while blue switch and blue pedal control
coag.
When cut is activated, the LED for cut startup will illuminate and the sound
with 750HZ will be sent; when coag is activated, the LED for coag startup will
illuminate and the sound with 250HZ will be sent. The hand control pencil and
foot control pencil can not be used at the same time because they share one
output channel.
Stop Operating