Operation manual
the floated level of the patient until normal output power is recovered. If you don’t
know the normal output power, please try from the low level. Make sure that output
level of machine start and end is low to prevent any risk to the patient causing by
high output power.
● The patients can not come in touch with the metal parts (such as bracket for operating
table), so the antistatic diaphragm is recommended. The medical staff must be
equipped with the rubber gloves with good insulation property in the operation to
prevent burns.
● The patient can come in touch with the metal by disinfectant, flushing fluid and
patients’ body fluid, causing burns. Therefore, the bed mat, matters and patients
should be kept dry.
● If the high frequency current might flow through the part of body with small cross
section in the operation, the bipolar technique (mode) is recommended to prevent the
undesired coagulation.
● When the high frequency surgical unit and physiological custodial unit are
simultaneously acted on one patient, the other custody electrodes should be kept
away from the operating electrodes and neutral electrodes, and you would better not
use the needle-shaped electrode. Under all circumstances, it is recommended to use
the custodial system with the restraint device for high frequency current.
● To prevent the operation electrode and neutral electrode cable coming in contact with
patients or other wire, please place the electrodes which are not in use in insulated
container.
● If a normal high frequency surgical unit have failures or the output decreases under
normal work setting, the neutral electrodes may be misused or a poor contact may
be between the electrodes. Under this circumstance, you should check the operation
and connection before selecting the higher output power.
● To avoid the HF radiation, the operators must not wear metal glasses in endoscopic
operation. Otherwise the operator’s face may be burned.
● Pay more attentions to the patients with metal in their bodies. To avoid burns, the HF
currents channel should be kept away from the metal in patients’ bodies for this may
cause the damage of their tissue.
● Do not touch the active electrode patient plate or electrode by the naked part of your
body for there is HF radiation on both electrode and patient plate, and the HF
radiation may cause pain and burns to the body.
● The accessories, including hand control pencil, foot control pencil and patient plate,
are consumable components which need to be check for working condition before
operating to prevent burns caused by false operation.
● Precautions to the burns of High frequency patient plate