4.14 Foot Hilow Motor
Chapter 4: Removal, Replacement, and Adjustment Procedures
15. Tighten the hoist (K) enough to remove pressure from the mounting pins
on the foot hilow drive.
16. Remove the rue ring (N) from the pin (O) on the motor end of the foot
hilow drive (P)
(see figure 4-21 on page 4-45).
17. Remove the nut (R) from the bolt (Q) on the shaft end of the foot hilow
drive (P).
18. While supporting the drive (P), remove the pin (O) from the motor end.
19. While supporting the drive (P), remove the bolt (Q) from the shaft end.
20. Remove the drive (P) from the bed.
Replacement
1. Install the motor end of the new motor into the bed.
2. Connect the motor cable to the motor control P.C. board.
3. Plug the bed into an appropriate power source.
4. Operate the hilow function as necessary to align the shaft end of the motor
with the lift arm on the bed.
5. Do the removal procedure in reverse order.
6. To make sure the VersaCare™ Bed operates correctly, do the “Function
Checks” on page 2-6.
Page 4 - 44 VersaCare™ Bed Service Manual (MAN333 REV 2)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hill-Rom VersaCare P3200 and is the answer not in the manual?

Hill-Rom VersaCare P3200 Specifications

General IconGeneral
TypeHospital Bed
Trendelenburg PositionYes
Reverse Trendelenburg PositionYes
Side RailsYes, integrated
Power SourceElectric
Wheel TypeLocking casters
Height Adjustment15" - 30"
Maximum Head Elevation70 degrees
Maximum Knee Elevation40 degrees
Height Adjustability15" - 30"
Head and Foot AdjustmentYes
Weight Capacity500 lbs

Related product manuals