EC08
Signal saturationThe amplitude of the
signal is too large due to activity or other
reason
Keep quiet, Press “start/stop button
to measure again.
EC09
overtimemeasuring time beyond 120s in
adult/child mode, 90s in neonatal mode.
Keep quiet, Press “start/stop button
to measure again.
EC10 manual de-activation
Keep quiet, Press “start/stop button
to measure again.
EC11 System error
Restart If the error occur frequently,
professional Maintenance may be
necessary
EC16
Cuff pressure beyond the max pressure
setting
Reprogramming in PC software and
up-regulate the max inflation
pressure setting
EC32 The handshake communication fail
Press “start/stop button to measure
again.
EC33
safe pressure beyond 15mmHg
immeasurable
Gassing and remeasure when the
cuff pressure decline to below
15mmHg
EC34
Measurement finish ,and restart the
measuring when the pressure is still above
15mmHg
Gassing and remeasure when the
cuff pressure decline to below
15mmHg
EC35
No answer when the measurement
command is sent.
Press “start/stop button to measure
again.
EC36 Measurement result is not available
Press “start/stop button to measure
again.
EC37 Overtime (beyond 180s) remeasure
EC48 Memory is full, can no longer measure Upload and empty the data
* If the errors can be revised, please call distributor or after-sale department of Hingmed
for help.
5.2 Maintenance
5.2.1 Maintenance inspection and safety management
Visually inspect the broken or damages on enclosure and cuff. If any damage is found,
do not use the monitor. Please contact the distributor or the after-sale service department
of Hingmed.
5.2.2 Host Maintenance
After use, it is important to perform preventative maintenance to ensure the safe and
efficient operation of the monitor for long-term.
It is recommended that the performance should be checked every 2 years and after
maintenance and repair, by retesting at least the requirements in limits of the error of the
cuff pressure indication and Air leakage (testing at least at 50 mmHg and 200 mmHg).
Do not disinfect. DO NOT immerse the monitor in any fluid, or attempt to clean with any
liquid detergents, cleaning agents, or solvents. You may use a soft, damp cloth to remove
dirt and dust from the monitor. If the unit does become immersed in water, do not use
and contact the distributor or our service department.