Power Measurement
11
1
2
3
4
5
6
7
8
9
10
11
4. Connect the Voltage cord to the voltage input terminals of
the 3169.
Connect the voltage cords to the instrument first, and then
to the active lines to be measured.
Observe the following to avoid electric shock and short cir-
cuits.
Voltage cord should only be connected to the secondary
side of a
breaker, so the breaker can prevent an accident if
a short circuit occurs. Connections should never be made
to the primary side of a breaker, because unrestricted cur-
rent flow could cause a serious accident if a short circuit
occu
rs.
Do not allow the voltage cord clips to touch two wires at the
same t
ime.
Never touch the edge of the metal clips.
Voltage input terminals U
1
, U
2
, and U
3
are common to the N
terminal and are not insulated. To avoid the risk of electric
shock, do not touch the terminals.
To prevent an electric shock accident, confirm that the
w
hite or red portion (insulation layer) inside the cable of
voltage cord is not exposed. If a color inside the cable is
exposed, do not use the cable.
For safety reasons, when taking measurements, only use the
L9438-53 Voltage Cord provided with the instrument.
Insert plugs all the way in.
Voltage input terminals
NU
1
U
2
Black Red Yellow