2.3 Current Measurement
mA
Measurement
(
40 mA/400 mA
)
Move the function selector to th
mA position.
2. Select DC ( ) or AC ( ) usin
the REC•MEMO button.
3. Connect the red test lead to ter
minal μA.mA, and the black tes
lead to terminal COM.
4. Connect the test leads to th
measurement object, and rea
the indicated value.
REC•MEMO
Black
Red
A
Measurement
(10 A range)
Note that the product may be damaged if current
exceeding the selected measurement range is
applied for a long time (for the 10 A range, continu-
ous current must be limited to 7A, or to less than
one minute if over 7A)
REC•MEMO
Black
Red
Move the function selector to th
10A position.
2. Select DC ( ) or AC ( ) usin
the REC•MEMO button.
3. Connect the red test lead to term
nal A, and the black test lead t
terminal COM.
4. Connect the test leads to the mea
surement object, and read th
indicated value.
The Auto Range function is no
provided.