2.3 Current Measurement
mA
Measurement
(
40 mA/400 mA
)
1.
Move the function selector to th
e
mA position.
2. Select DC ( ) or AC ( ) usin
g
the REC•MEMO button.
3. Connect the red test lead to ter
-
minal μA.mA, and the black tes
t
lead to terminal COM.
4. Connect the test leads to th
e
measurement object, and rea
d
the indicated value.
REC•MEMO
Black
Red
A
Measurement
(10 A range)
Note that the product may be damaged if current
exceeding the selected measurement range is
applied for a long time (for the 10 A range, continu-
ous current must be limited to 7A, or to less than
one minute if over 7A)
REC•MEMO
Black
Red
1.
Move the function selector to th
e
10A position.
2. Select DC ( ) or AC ( ) usin
g
the REC•MEMO button.
3. Connect the red test lead to term
i-
nal A, and the black test lead t
o
terminal COM.
4. Connect the test leads to the mea
-
surement object, and read th
e
indicated value.
The Auto Range function is no
t
provided.