29
3.4 Relative Display Function
Canceling the Relative
function:
Press H.AUTO/REL
1. Move the function selecto
r
to the Ω position and con
-
nect the test leads to th
e
equipment.
2. Short-circuit the test lead
s.
The resistance of the te
st
leads is displayed.
3. Press the H.AUTO/RE
L
button for at least 1 se
c-
ond.
("REL" lights up, "0000")
4. Connect the test leads
to
the measurement objec
t,
and read the indicate
d
value.
Application: Using the Relative Display function for the purpose
o
f zero adjustment in the measurement of low resistances
H.AUTO/REL
Measure the resistance.
for at least
1 second
Short-circuit the test leads