3.4 Relative Display Function
Canceling the Relative
function:
Press H.AUTO/REL
1. Move the function selecto
to the Ω position and con
nect the test leads to th
equipment.
2. Short-circuit the test lead
The resistance of the te
leads is displayed.
3. Press the H.AUTO/RE
button for at least 1 se
ond.
("REL" lights up, "0000")
4. Connect the test leads
the measurement objec
and read the indicate
value.
Application: Using the Relative Display function for the purpose
o
f zero adjustment in the measurement of low resistances
H.AUTO/REL
Measure the resistance.
for at least
1 second
Short-circuit the test leads