47
Chapter 3 Specifications
Location for use Indoor, altitude up to 2000 m (6562 feet)
Applicable standards Safety:
EN 61010-1:1992+A2:1995
CAT III (expected transient overvoltage:
6000 V), Pollution level 2,
EN 61010-2-031:1994
EN 61010-2-032:1995
EN 60529:1991 IP40 (protected against
access to hazardous parts with a wire)
EMC:
EN 61326-1:1997+A1:1998
Maximum conductor
diameter for
measurement
φ 55 mm max.
Operating temperature
and humidity range
0to40 (32 to 104 ),
80%RH or less (no condensation)
Temperature
characteristics
Current and voltage
measurement
In 0 to 40 range: 0.1 x accuracy
specifications/
Phase detection
circuit
In 0 to 40 range: Within 2deg.
Storage temperature
range
-10 to 50
(14 to 122 , no condensation)
Power source 6LR61, 6LF22 alkaline battery
9V x 1
Output function Optical insulation output
(using optional 9636 RS-232C cable)
Maximum power
consumption
220 mVA
Battery life Alkaline battery (6LR61, 6LF22)
approx. 25 hours
Manganese battery (6F22) approx. 10 hours
External dimensions 62(W) x 260(H) x 39(D) mm
2.44"(W) x 10.24"(H) x 1.54"(D)
Mass Approx. 550 g
Approx. 19.4 oz.