20
Chapter 2 Measurement Procedure
I/ key
U/
key
AUTOrange 150Vrange 300Vrange 600Vrange
AUTOrange 200Arange 1000Arange
Table 1.
Range Composition for Single-Phase Power Measurement
U Voltage range
I 150.0V 300.0V 600V
Current
range
200.0A 30.00k 60.00k 120.0k
1000A 150.0k 300.0k 600.0k
Unit [W] or [VA] or [var]
2.3 Range Setup
1. Press the
RANGE
key. The voltage range then
appears on Display 2 and current range on Display
3. In this condition, Display 2 and Display 3
should be blinking.
2. To change the voltage range, press the
U/
key.
To change the current range, press the
I/
key.
The power range varies with the combination of
voltage and current ranges as listed in Tables 1 and
2.