8.4 Communication Methods
118
*1: Only the MAV bit (bit 4) is cleared.
*2: All bits except the MAV bit are cleared.
*3: Except the PON bit (bit 7).
During communications, REMOTE is lit to indicate the remote control state.
To cancel the
Remote state
Initialization Items
: initialized,
: not initialized
Initialization Method
Item
At Power-
on
RST
Command
Device
Clear
CLS
Command
Device-specific functions
(Range, etc.)
Output Queue
Input buffer
Status Byte Register
*1
*2
Event registers
*3
Enable register
Current path
Headers on/off
Local Function
REMOTE off
• Remote control can be canceled by pressing the SHIFT key and then the
AUTO key.
• If the Local Lock Out (p.109) GP-IB command has been issued, the
Remote state cannot be canceled.