8.6 Message Reference
143
8
Chapter 8 RS-232C/GP-IB/LAN Interfaces
Query the Maximum value
(Resistance Measurement)
Syntax
Query
:CALCulate:STATistics:RESistance:MAXimum?
Response <Maximum value (NR3)>,<Data No. of Maximum value (NR1)>
Example
Query
:CALC:STAT:RES:MAX?
Response
297.28E-3,15
(Voltage Measurement)
Syntax
Query
:CALCulate:STATistics:VOLTage:MAXimum?
Response <Maximum value (NR3)>,<Data No. of Maximum value (NR1)>
Example
Query
:CALC:STAT:VOLT:MAX?
Response
1.3924E+0,1
Query the Minimum value
(Resistance Measurement)
Syntax
Query
:CALCulate:STATistics:RESistance:MINimum?
Response <Minimum value (NR3)>,<Data No. of Minimum value (NR1)>
Example
Query
:CALC:STAT:RES:MIN?
Response
294.88E-3,8
(Voltage Measurement)
Syntax
Query
:CALCulate:STATistics:VOLTage:MINimum?
Response <Minimum value (NR3)>,<Data No. of Minimum value (NR1)>
Example
Query
:CALC:STAT:VOLT:MIN?
Response
1.3923E+0,2