4.1 Comparator Function
46
12
Press so that the indicated position blinks, and select the comparison method for
the comparator (here, HIGH/LOW).
HIGH, LOW ..... Compare by upper and lower thresholds (default setting)
REF, %............. Compare by reference value and tolerance
13
Switch to the upper/lower threshold setting display, and specify the thresholds.
14
Applies setting and returns to the Measurement display.
The comparator function is enabled.
To cancel the settings: SHIFT key
15
Connect a test object and judge the measured value.
In the V mode, you can verify comparator settings by pressing the VIEW key.
See " Switching Between Measurement Value and Comparator Setting Displays" (p.58).
HIGH & LOW flashing
Or
numeric key-
pads
For this example,
Upper Threshold: 15.2 V
Upper Threshold: 15 V
COMP lit
Measured Resistance
Judgment Result
Measured Voltage
Upper Threshold Value < Measured Value
Lower Threshold Value
Measured Value Upper Threshold Value
Measured Value < Lower Threshold Value