7
Making Measurements
Measuring load current
Precautions for measuring load current
IMPORTANT
Clamp the instrument around only one conductor wire. The instrument cannot measure load
current when clamped around two or more wires together in a bundle, regardless of the single-
phase and three-phase.
Put the conductor perpendicular to the sensor.
The instrument may not be able to measure rush current or signicantly uctuating current
correctly.
The instrument can display a readout of other than zero without input at a low temperature;
however, this does not a󰀨ect any measurement.
OK
NO
NO
The instrument may not be able to measure special waveforms, such as those
owing through the secondary side of an inverter.
Depending on the magnitude of the input current, the jaws may generate a sound
due to resonance; however, this does not a󰀨ect any measurement.
When you cannot nd the magnitude of an input current, disable the lter function
and start measurement using the auto range or 600.0 A range.
HIOKI CM4001A963-00