57
Accuracy Table
4
Diode
Range Accuracy
Measurement
current
Open circuit
voltage
1.000 V ±10.0% rdg. 0.5 mA
3.0 V DC or less
Voltage drop
due to battery
consumption
• Overload protection: 1000 V DC/1000 V AC or 10
6
V • Hz (energized for 1
minute)
Current under short circuit: 0.7 mA or less
Current under overload: Steady state 15 mA or less, transient state 0.8 A or
less
5
Continuity
Range Accuracy
Measurement
current
Open circuit
voltage
400.0
±1.0% rdg. ±15 dgt. Approx. 140 A 0.5 V DC or less
• Overload protection: 1000 V DC/1000 V AC or 10
6
V • Hz (energized for 1
minute)
Current under short circuit: 300 A or less
Current under overload: Steady state 15 mA or less, transient state 0.8 A or
less
• Continuity ON threshold: 90 ± 40
or less (buzzer)
• Response time: Open circuit or short circuit is detected for at least 0.5 ms.