58
Accuracy Table
6
Electrostatic capacity
Range Accuracy Charging current Open circuit voltage
50.00 nF ±1.5% rdg. ±15 dgt.
Approx. 30 A 1.5 V DC or less
500.0 nF ±2.0% rdg. ±5 dgt.
5.000 F
±5.0% rdg. ±5 dgt. 50.00 F
100.0 F
• Overload protection: 1000 V DC/1000 V AC or 10
6
V • Hz (energized for 1
minute)
Current under short circuit: 50 A or less
Current under overload: Steady state 15 mA or less, transient state 0.8 A or
less
• Accuracy guarantee condition: After REL (zero adjustment) has been
performed
7
Frequency
Range Accuracy
Minimum sensitivity
voltage
5.000 Hz
±0.1% rdg. ±3 dgt.
Square wave of 1.5 Vrms
or more
50.00 Hz
500.0 Hz
5.000 kHz
50.00 kHz
500.0 kHz
5.000 MHz ±0.1% rdg. ±3 dgt.
Square wave of 2.0 Vrms
or more
• Measurement range: 1 Hz or more