Loading...
34
*1: Display value before zero adjustment
*2: Accuracy is applicable after zero adjustment (when the temperature
changes more than 1°C, zero adjustment is necessary)
Resistance Measurement
Open-circuit
voltage
4.0
to 6.9 V
Measuring
current
200 mA or more (at 6 or less
*1
)
Effect of
temperature
*2
±3%rdg.±2dgt.
(applicable to the operating temperature range
other than 18C to 28C.)
Effect of supply
voltage
*2
±3%rdg.±2dgt. and within allowance
Response time Within 1 s (measurement terminal open short)
Possible
number of
measurements
200 times or more
Overload
protection
600 VAC (10 s,
by Fuse
)
Zero adjustment
range
0 to 3
Display update
interval
Within 1 s
Display
range
(Auto range)
Maximum
indicated value
Resolution
Accuracy
*2
(after zero adjustment)
10 10.00 0.01
0 to 0.19
: ±3dgt.
0.20 to 10.00
: ±3%rdg.±2dgt.
100 100.0 0.1
±3%rdg.±2dgt.
1000 1000 1
Range configuration