8
*1: Pulsating currents with an overlapping AC component of 30 V or
greater are detected as AC.
*2: Applicable to the operating temperature range other than 18 to 28C.
*3: Minimum indicated value: 30.0 V
*4: Ranges in excess of 600 V are outside the accuracy guarantee.
Voltage Measurement
AC/DC
automatic
detection range
AC detected at 30 V or greater (50/60 Hz)
*1
Effect of
temperature
Measurement accuracy per 1°C
× 0.1
*2
Overload
protection
750 VAC (10 s), 750 VDC (10 s)
Display update
interval
Within 1 s
Response time
Within 1.2 s
(when input voltage is cycled from 0 V to 600 V)
Input
resistance
100 k or more (50/60 Hz)
Frequency
range
50/60 Hz
Range configuration
Display range
(Auto range)
Maximum
indicated value
Resolution Accuracy
420 V
*3
420.0 V 0.1 V
±2.3%rdg.±8dgt
*4
600 V 750 V 1 V
Input
resistance
100 k
or more
Range configuration
Display range
(Auto range)
Maximum
indicated value
Resolution Accuracy
4.2 V 4.200 V 0.001 V
±1.3%rdg.±4dgt
*4
42 V 42.00 V 0.01 V
420 V 420.0 V 0.1 V
600 V 750 V 1 V
AC Voltage Measurement
DC Voltage Measurement