6 - 10
6-1-8
Creating Reference Data
Start the setup software.
For subsequent operation, operate the PC
mouse while watching the PC monitor screen.
Information:
It takes approx. 20 minutes to create reference
data.
The following shows the times required for
each item:
Matrix: Approx. 10 minutes
White Balance: Approx. 10 minutes
White Balance Abs: Approx. 30 seconds
Preparation:
1) Point at light box without chart, fi lling the
screen.
2) Prepare the C12 light balancing fi lter (step-
up rings): Attach it during setup.
Procedure:
1) Choose MATRIX.
2) Click the EXECUTE button.
Adjustment > Creating Reference Data
3) The ATTACH THE FILTER dialog will
appear during setup. Attach the C12 light
balancing fi lter over the lens of DVD video
camera/recorder, and then click the OK
button.
4) The REMOVE THE FILTER dialog will
appear during setup.
Remove the C12 light balancing fi lter from
the lens of DVD video camera/recorder, and
then click the OK button.
After that, the ATTACH THE FILTER dialog
and REMOVE THE FILTER dialog may
occasionally appear: Reattach the C12 light
balancing fi lter and remove it each time.
5) When setup is complete, the SETUP
FINISHED dialog will appear: Click the OK
button to restore the SETUP MENU screen.
When setup is complete with the C12 light
balancing fi lter attached, remove the C12
light balancing fi lter.
SETUP MENU screen.
MANUAL ADJUSTMENT PROGRAM for SERVICE STATION
Matrix
EXECUTE
RETURN
SETUP MENU
White Balance Abs
White Balance
2
1
ATTACH THE FILTER dialog
REMOVE THE FILTER dialog
OK
Attach The Filter.
CHECK
!
3
OK
Remove The Filter.
CHECK
!
4
SETUP FINISHED dialog
FINISHED
SETUP FINISHED.
OK
5