78 Warnings
Mask warning
If the mask slips off your face, the tube is pulled out or any other leak occurs that can-
not be compensated for, the device does not switch off automatically but emits an
acoustic and visual warning.
The mask warning is only active in the „Auto OFF“ and „Auto Start“ automatic modes.
Testing the mask warning
Start the device in „Auto OFF“ mode. The mask should be open, not on the face. After
a short time an acoustic signal sounds.
Pressure warning
Should an excessively high pressure occur during the treatment due to a hard-
ware error or other circumstance, the device emits an signal sound and switches
the blower off.
The warning is turned off by pressing the ON/OFF key.
Alarm clock
Please note: Ex factory, the alarm clock function is not enabled. But it can be unlo-
cked by the physician or dealer with TRENDset PC software.
If enabled, you can activate or deactivate the alarm clock with the selection keys
and . Pressing the ON/OFF key once mutes the alarm for the next 5 minutes and
pressing it twice stops it completely.