Manufacturer’s declaration on electromagnetic
compatibility
Point 2 satisfies the IEC 60601-1-2:2014 standard and is intended for use in the elec-
tromagnetic environment described below. Deviating ambient conditions can impair
the essential performance features such as pressure accuracy and alarm, or lead to
the breakdown of the device.
Guidance and manufacturer’s declaration – electromagnetic emissions
The point 2 ventilator is intended for use in the electromagnetic environment specified below.
The user of the point 2 ventilator should ensure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions acc. to CISPR
11
Group 1 The point 2 ventilator uses RF energy only for
its internal function. Therefore, RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions acc. to CISPR
11
Class B
The point 2 ventilator is suitable for use in
all establishments including those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/flicker emissions
acc. to IEC 61000-3-3
complies