Annex 93
Setting ranges and control accuracy
Parameter Settings range
Dependency Setting steps
Language
DEU, ENG, ELL, SPA, FRA, ITA,
NLD, TUR, PLK, POR, CZE
- -
P-Unit hPa, mbar, cmH
2
O
-
-
Press.
1
4 – 20 (25
7
) hPa - 0.5 hPa
IPAP
2
EPAP – 20 (25
7
) hPa
[≥ EPAP, ≤ 20 (25
7
) hPa]
≥ EPAP, ≤ 20 (25
7
) hPa
0.5 hPa
EPAP
2
4 hPa – IPAP
[≥ 4 hPa, ≤ IPAP]
≥ 4 hPa, ≤ IPAP 0.5 hPa
I-Slope
2
1 – 3
[IPAP, EPAP, frequency, I:E]
IPAP, EPAP,
Frequency, I:E
1 Level
E-Slope
2
1 – 3
[IPAP, EPAP, frequency, I:E]
IPAP, EPAP,
Frequency, I:E
1 Level
Frequency
3
5 – 30 bpm - 1 bpm
I:E
3
20 – 80 %
-
5%
Delay Time
4
0 – 20 s - 1 s
Backup freq
5
ON, OFF - -
P-Min
6
4 – 20 hPa
[≤ P-Start]
≤ P-Start 0.5 hPa
P-Max
6
4 – 20 hPa
[≤ P-Start]
≥ P-Start 0.5 hPa
P-Start
6
4 – 20 hPa
[≥ P-Min, ≤ P-Max]
≥ P-Min,
≤ P-Max
0.5 hPa
I-FLEX 0 – 3 hPa/ (l/s) - 1 hPa/ (l/s)
E-FLEX 0 – 3 hPa/ (l/s)
-
1 hPa/ (l/s)
Ramp
0 – 45 min - 5 min
P-Ramp 4 hPa – 20 (25
7
) hPa
CPAP: ≤ Press.
APAP
6
: ≤ P-Start
BILEVEL S, T, ST
2
:
≤ EPAP
0.5 hPa 0.5 hPa
1
only point 2 CPAP and point 2 AutoCPAP and point 2 BILEVEL ST20/25 in modes CPAP
2
only point 2 BILEVEL ST20/25
3
only point 2 BILEVEL ST20/25 in modes BILEVEL T and ST
4
only point 2 BILEVEL ST20/25 in BILEVEL ST mode
5
only point 2 BILEVEL ST20/25 in BILEVEL S mode
6
only point 2 AutoCPAP
7
only point 2 BILEVEL ST25