34 Commissioning
Using oxygen
Risk of fire and burns!
Oxygen sources should be at a distance of more than 1 m
from the device in order to avoid the risk of fires and burns.
Please note:
Before using oxygen, the safety instructions as of page 23 must be read.
The use of oxygen affects the therapeutic pressure. For this reason, we recommend
the use of oxygen already during the titration process.
If the device is in standby mode or switched off, the oxygen supply must always
be switched off.
When supplying oxygen, please use a kink-resistant tube made of a medically appro-
ved material. In order to prevent back flow of oxygen into the device, we recommend
the use of an O
2
pressure valve.
Supplying oxygen directly in the mask
Mask
Therapy
tube
O
2
pressure valve
O