84 Cleaning and Disinfection
Use of the device by several patients
If the device is intended for use by more than one patient (e.g., when operated in clin-
ics), a suitable bacterial filter (e.g., Medisize Barr-vent S) must be used continuously
to protect the device from contamination by human pathogens. The bacterial filter
must be changed daily.
Risk of infection from germs!
•
If the device is not continuously protected by a bacterial
filter, there is a risk of it being contaminated with human
pathogens. If the device is used on another patient, the
patient can be infected with these germs.
Ö Protect the device continuously with a bacterial filter.
Ö For each new patient clean and disinfect the device so
that it is free from human pathogens.
Ö
If there is any doubt, always assume that the device is
contaminated, and hygienically prepare it according to
the hygiene concept.
Ö
If contamination with multiresistant pathogens (e.g.
MRSA) is suspected, the device must be packaged, with
the appropriate labeling, and disinfected accordingly.
• Accessories and consumables (e.g., tubing, mask etc.) may
be contaminated with human pathogens. If these are used
on another patient, the patient can be infected with these
germs.
Ö
Accessories and consumables designated for "single use”
must only be used once on one patient.
Ö
Accessories and consumables designated for "single
patient use" must only be used on one patient and
must be cleaned as per the manufacturer's instructions.