AreaRAE Plus & AreaRAE Pro User’s Guide
101
Important! When you are done setting thresholds, exit Diagnostic Mode and test the instrument before
actual use.
19.4.4 Setting Pump Stall Threshold Values Static Method
Use the following values for reference when using the Static method to adjust the pump stall threshold
values:
Low Speed
High Speed
Vacuum
-2.5in Hg
-10in Hg
Flow rate
>200 cc/min
>400 cc/min
Idle (I)
100 to 200 counts
150 to 250 counts
Block
(Block-Idle) > 100 counts
(Block-idle) > 100 counts
Stall Setting
(Idle + Block)/2
(Idle + Block)/2
Stall High Threshold Setting Static Method
In Diagnostic Mode, press the [MODE] key repeatedly until the Pump screen is displayed. High”
should be highlighted. Otherwise, press [Y/+] to select the High value.
The Idle value (shown as a value for “I”) should be 150 to 250. Record the value.
Block the inlet, and record the value after it goes up.
The blocked value minus the Idle (Unblocked) value should be greater than 100 counts: (Blocked
value Unblocked value) > 100.
If it is greater than 100, then the pump is working correctly and tubing leading from the inlet is not
leaking.
To calculate the best High value, first add the Blocked and Unblocked values and divide by 2:
(Blocked value + Unblocked value) / 2 = correct Stall High value.
Press [N/-] to advance to the next Pump screen.
Use the [Y/+] and [N/-] keys to increase or decrease the high value input the results of your
calculation.
Press [MODE] to register that value.