Problem Cause Solution
Detector Doesn’t
accurately measure gas.
Sensor(s) require
calibration.
Carry out calibration.
Detector is colder/hotter
than gas temperature.
Allow the Detector to attain
ambient temperature before
use.
The sensor filter is blocked. Replace sensor filter
The detector does not
alarm.
Alarm setpoints set
incorrectly.
Define the alarm setpoint in
Detector Configurator.
Alarm setpoints set to zero.
Define the alarm setpoint in
Detector Configurator.
Detector is in calibration
mode.
Complete the calibration
procedure.
Detector is in DC mode.
Stop data communication via
a mobile phone.
Detector is in IR
communication.
Stop data communication via
IR Link.
The Detector alarms
without reason
Thesensor was exposed to
the target gas.
Detector is operating
normally. Use caution in
suspected areas. Check the
peak gas exposure reading.
Alarm setpoints are set
incorrectly.
Define the alarm setpoint in
Detector Configurator.
Sensors require calibration. Carry out calibration.
Missing or faulty sensors. Replace the sensors.
Battery temperature is out
of acceptable range.
Move to lower temperature
ambient to charge the battery.
Battery indicator doesn’t
display when charging.
Battery is depleted.
Charge the battery for 8 hours.
If the battery indicator doesn’t
light after charging, contact
Honeywell
BW Flex Series 45 User Manual

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell BW Flex Series and is the answer not in the manual?

Honeywell BW Flex Series Specifications

General IconGeneral
BrandHoneywell
ModelBW Flex Series
CategoryGas Detectors
LanguageEnglish

Related product manuals