For possible sensor installation, refer to this table:
Slot 1 Slot 2 Slot 3 Slot 4 Slot 5 Slot 6
MIPEX NDIR Sensor ●
Dynament NDIR Sensor ● ● ● ●
PID Sensor ●
LEL Sensor ●
EC Sensor ● ● ● ● ●
●Sensor locations
EC Sensor: Max.5pcs
MIPEX NDIR Sensor: Max.1pc
PID Sensor: Max.1pc
LEL Sensor: Max.1pc
Dynament NDIR Sensor: Max.3pcs
In addition, there is a noise sensor located inside of the BW RigRat, which can measure the
ambient noise. The BW RigRat also contains a THP sensor in pumped version that can measure
the ambient temperature, humidity, and gas flow outside.
Electrical Parameters
The BW RigRat provides 5 external connectors with the following electrical parameters:
l
AC charger connector (use only in a non-hazardous area): Um=6V
l
IS charger connector: Ui=24V, Ii=150mA, Pi=1.15W, Ci=0.36nF, Li=0 (for all);
Ui=12V, Ii=500mA, Pi=1.15W, Ci=0.36nF, Li=0 (for NA
certification)
l
ON-OFF switch connector: Ui=30V, Ii=100mA, Pi=0.75W, Ci=1.1nF, Li=0
l
4~20mA input connector: Ui=30V, Ii=100mA, Pi=0.75W, Ci=0, Li=0
BW RigRat 16 User Manual