Type Code Display message Key Action Note
1006
TH Module Not
Found
Click to
acknowledge
Power off
in 300s
Reinstall TH Module
1007
Datalog Flash Not
Found
Click to
acknowledge
Power off
in 300s
Datalog memory broken or missing
1008
Bluetooth (BLE) not
Found
Click to
acknowledge
Power off
in 300s
Bluetooth (BLE) module not found
1009 GPS Not Found
Click to
acknowledge
Power off
in 300s
GPS module not found
1010
Wireless Module Not
Match Configuration
Click to
acknowledge
Power off
in 300s
Invalid wireless configuration
1011
Mesh Module Not
Found
Click to
acknowledge
Power off
in 300s
1013
Wi-Fi Module Not
Found
Click to
acknowledge
Power off
in 300s
1017
Battery Charging
Current Too Low
Click to
acknowledge
Power off
in 300s
1018 Battery NTC Fault
Click to
acknowledge
Power off
in 300s
Temperature out of range, battery
cannot charge
1020 Forced to charge
Click to
acknowledge
Power off
in 300s
Battery voltage too low, power on
with AC
1022 RAEMet Not Found
Click to
acknowledge
Power off
in 300s
RAEMet sensor not found. Reinstall
RAEMet to Serial receptacle.