Reset TWA/STEL and/or Peak Readings
Before you begin.
You need to enable TWA/STEL Reset and/or Peak Reset in Safety Suite Device Configurator
(SSDC) to reset readings in the detector.
1. Open Safety Suite Device Configurator (SSDC) in your PC.
2. Login using your username (Default: administrator) and password (Default: Default123)
3. Ensure the device is connected to the computer and powered on.
4. Click on the serial number:
5. Go to the Settings tab and scroll down to "Instrument user preferences"
6. Click on the dropdown menu from "Clear Peaks" and choose which readings to reset.
7.
Click on the "Save" button to save your settings to the device.
Zero Sensors
Before you begin.
Connect the Nitrogen if this is a CO
2
unit.
1. Go to the main menu and select Zero Sensors.
2. Press and hold for 3 seconds. A Zeroing process starts automatically. The screen displays
all current gas measurements, highlighting entries above zero. Ambient air is applied to
zero all sensors that are not CO
2
. The screen displays all current gas measurements,
highlighting entries being reset to zero.
BW Ultra 24 User Manual