บริษัท เอดีดี เฟอร์เนส จํากัด 
ADD FURNACE CO.,LTD. 
44 ซอยบรมราชชนนี 70 ถนนบรมราชชนนี แขวงศาลาธรรมสพน์ เขตทวีวัฒนา กรุงเทพฯ 10170 
โทร: 02-888-3472 โทร: ออกแบบ:08-08-170-170 แฟกซ์: 02-888-3258 
https://www.add-furnace.com E-mail: sales@add-furnace.com 
 
 
   
SERVICE 
CAUTION 
Open  the  master  switch  to  disconnect  power 
before removing or installing the detector or its 
cover.  More  than  one  disconnects  can  be 
involved. 
 
Periodic Maintenance 
A 
Clean  the  viewing  window  (or  focusing  lens)  when 
necessary.  Remove  the  detector  (see  Troubleshooting 
section) and use a  clean  cloth  over  the  eraser end  of  
a pencil. Do not remove the window (or lens) to clean it. If 
it is broken or damaged or it is coated with a substance   
that cannot be cleaned off, replace it (see Fig. 14). 
B 
Keep the flame detection system adjusted for the 
smoothest,  most  reliable  operation  as  recommended 
by  the burner manufacturer. 
C 
Replace the sensing tube, coil, and shutter assembly, or 
viewing  window  only  when necessary,  to  obtain proper 
operation. 
 
Removing the Detector Cover 
Open the Master Switch 
C7061A:   Unscrew the four captive cover screws (Fig. 8   
and 9) and carefully slide off the cover. 
NOTE: These bolts are removable. Put them in a 
safe place to avoid losing them. 
C7061F  Carefully  unscrew  the cover from  the  detector
 
cap. 
 
 
 
 
Fig. 12. Replacing ultraviolet radiation sensing tube. 
 
Replacing the Coil and Shutter Assembly 
NOTE: Use only a 190971 B coil and shutter assembly. 
 
A 
Open the master switch and remove the detector cover 
(see Removing the Detector Cover section). 
B 
Remove the ultraviolet sensing tube (steps A through E 
of Replacing the Ultraviolet Sensing Tube section). 
Replacing the Ultraviolet Sensing Tube 
A 
Open the master switch and remove the cover from the 
detector (see instructions above). 
B 
Locate the UV sensing tube. 
 
IMPORTANT 
Be very careful not to kink or otherwise damage 
the flexible shutter. 
 
C 
Gently bend the alignment guide just enough to free the 
tip 
D 
of the tube.
 
Insert a screwdriver between the tube base and socket, 
E  
and gently pry the tube out of its socket.
 
F  
Pull the tube out of its socket.
 
Insert the new tube through the openings in the shutter  
G 
assembly.
 
Align the three pins on the new tube with the holes in the 
H 
socket.
 
Carefully  push  the  new tube firmly  into  its socket; the 
alignment guide will snap into place around the tip of the 
I   
tube.
 
J  
Make sure the new UV sensor tube is seated securely.
 
C 
Cut the white wires as close as possible to the crimped 
connectors and remove the crimped connectors. 
D 
Remove the three mounting screws from  the  base of    
the coil and shutter assembly. Put the screws in a safe 
place. 
E 
Remove the coil and shutter assembly. 
F 
Put the new coil and shutter assembly into place. 
 
IMPORTANT 
Be very careful not to kink or otherwise damage 
the flexible shutter. 
 
G Insert the three mounting screws into the base of the coil 
and shutter assembly and tighten securely. 
H 
Remove sufficient insulation from each of the two white 
lead wires remaining on the detector, and from each of 
the two white lead wires on the new coil. 
I 
Using solderless connectors, connect one of the coil 
wires to one of the remaining white lead   wires. 
Connect the other coil wire to the other remaining white 
J   lead wire. Reinstall the sensing tube (steps F through I 
of Replacing the Ultraviolet Sensing Tube section). 
K  Replace the detector cover.