EasyManuals Logo

Honeywell C7061A1020 User Manual

Honeywell C7061A1020
16 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #14 background imageLoading...
Page #14 background image
EN1C-0015 R2105
14
บรษัท เอดีดี เฟอรเนส จํากัด
ADD FURNACE CO.,LTD.
44 ซอยบรมราชชนน 70 ถนนบรมราชชนน แขวงศาลาธรรมสพน์ เขตทวีวฒนา กรงเทพฯ 10170
โทร: 02-888-3472 โทร: ออกแบบ:08-08-170-170 แฟกซ์: 02-888-3258
https://www.add-furnace.com E-mail: sales@add-furnace.com
SERVICE
CAUTION
Open the master switch to disconnect power
before removing or installing the detector or its
cover. More than one disconnects can be
involved.
Periodic Maintenance
A
Clean the viewing window (or focusing lens) when
necessary. Remove the detector (see Troubleshooting
section) and use a clean cloth over the eraser end of
a pencil. Do not remove the window (or lens) to clean it. If
it is broken or damaged or it is coated with a substance
that cannot be cleaned off, replace it (see Fig. 14).
B
Keep the flame detection system adjusted for the
smoothest, most reliable operation as recommended
by the burner manufacturer.
C
Replace the sensing tube, coil, and shutter assembly, or
viewing window only when necessary, to obtain proper
operation.
Removing the Detector Cover
Open the Master Switch
C7061A: Unscrew the four captive cover screws (Fig. 8
and 9) and carefully slide off the cover.
NOTE: These bolts are removable. Put them in a
safe place to avoid losing them.
C7061F Carefully unscrew the cover from the detector
cap.
Fig. 12. Replacing ultraviolet radiation sensing tube.
Replacing the Coil and Shutter Assembly
NOTE: Use only a 190971 B coil and shutter assembly.
A
Open the master switch and remove the detector cover
(see Removing the Detector Cover section).
B
Remove the ultraviolet sensing tube (steps A through E
of Replacing the Ultraviolet Sensing Tube section).
Replacing the Ultraviolet Sensing Tube
A
Open the master switch and remove the cover from the
detector (see instructions above).
B
Locate the UV sensing tube.
IMPORTANT
Be very careful not to kink or otherwise damage
the flexible shutter.
C
Gently bend the alignment guide just enough to free the
tip
D
of the tube.
Insert a screwdriver between the tube base and socket,
E
and gently pry the tube out of its socket.
F
Pull the tube out of its socket.
Insert the new tube through the openings in the shutter
G
assembly.
Align the three pins on the new tube with the holes in the
H
socket.
Carefully push the new tube firmly into its socket; the
alignment guide will snap into place around the tip of the
I
tube.
J
Make sure the new UV sensor tube is seated securely.
C
Cut the white wires as close as possible to the crimped
connectors and remove the crimped connectors.
D
Remove the three mounting screws from the base of
the coil and shutter assembly. Put the screws in a safe
place.
E
Remove the coil and shutter assembly.
F
Put the new coil and shutter assembly into place.
IMPORTANT
Be very careful not to kink or otherwise damage
the flexible shutter.
G Insert the three mounting screws into the base of the coil
and shutter assembly and tighten securely.
H
Remove sufficient insulation from each of the two white
lead wires remaining on the detector, and from each of
the two white lead wires on the new coil.
I
Using solderless connectors, connect one of the coil
wires to one of the remaining white lead wires.
Connect the other coil wire to the other remaining white
J lead wire. Reinstall the sensing tube (steps F through I
of Replacing the Ultraviolet Sensing Tube section).
K Replace the detector cover.

Other manuals for Honeywell C7061A1020

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell C7061A1020 and is the answer not in the manual?

Honeywell C7061A1020 Specifications

General IconGeneral
BrandHoneywell
ModelC7061A1020
CategorySecurity Sensors
LanguageEnglish

Related product manuals